Abstract
Spectrophotometric measurements of the Fricke–gelatin–xylenol orange (FGX) gel dosimeter demonstrated reproducible linear dose response up to 25 Gy. However, oxidation processes continue post-irradiation, affecting the response of this dosimetry system. Additional oxygenation during preparation increases the sensitivity of the gel but does not improve the auto-oxidation stability of the dosimeter post-irradiation. A suitably stable gel composition that is recommended for radiotherapy dosimetry measurements contains 0.5 mM ferrous ammonium sulphate, 50 mM sulphuric acid, 0.15 mM xylenol orange and 3.0% by weight gelatin.