Article ID Journal Published Year Pages File Type
1967858 Clinica Chimica Acta 2007 8 Pages PDF
Abstract

BackgroundMeasurement of thyroid peroxidase autoantibodies (TPOAb) is useful in diagnosing patients with autoimmune thyroid disease. Measurement of thyroglobulin autoantibodies (TgAb) is used to detect potential interferences with thyroglobulin immunoassays and in limited situations for the diagnosis of autoimmune thyroid disease.MethodsThe limit of detection, imprecision, reference interval, method comparison and diagnostic concordance for the ADVIA Centaur, ARCHITECT i2000, AxSYM, Immulite 2000, Modular E170 (TPOAb only), and UniCel DxI 800 (TgAb only) methods were evaluated. The Advantage was used as the comparison method.ResultsTotal imprecision ranged from 2.6% to 14.9% for TgAb and 2.1% to 15.8% for TPOAb. Passing–Bablok slopes ranged from 0.51 to 10.4 (TgAb) and 1.05 to 7.12 (TPOAb) with correlation coefficients of 0.48 to 0.82 (TgAb) and 0.66 to 0.78 (TPOAb). Assay cutoffs were adjusted using a common set of reference interval samples. Concordance with the Advantage assay using the new cutoffs was found to be improved and ranged from 68.5% to 84.7% (TgAb) and 77.5% to 84.7% (TPOAb).ConclusionsAlthough all assays generally performed well, assay concordance for a negative or positive result ranged from 54.2 to 84.7%. Quantitative agreement between methods was generally poor and methods could not be used interchangeably. Additional standardization efforts are required to improve inter-method agreement.

Related Topics
Life Sciences Biochemistry, Genetics and Molecular Biology Biochemistry