ObjectivesEvaluation of automated flow cytometric analysis of white blood cell (WBC) count in peritoneal fluids.MethodsOne hundred peritoneal fluids were analyzed with manual microscopy, Sysmex XE-2100 and XE-5000, Siemens Advia 2120, Mindray BC-6800, Abbott Sapphire.ResultsHigh correlations (0.978 to 0.999) and modes biases (− 132 to 80 WBC/mm3) were found. Agreement at septic peritonitis cutoff ranged between 96% and 99%.ConclusionsThese hemocytometers display acceptable performance for WBC screening in peritoneal fluids.
► Correlations with manual microscopy were between 0.978 and 0.999. ► The mean bias comprised between − 132 and 80 WBC/mm3. ► The agreement at diagnostic cutoff of septic peritonitis was between 96 and 99%. ► The instruments had acceptable performance for routine peritoneal fluid screening.