Article ID Journal Published Year Pages File Type
1969842 Clinical Biochemistry 2011 6 Pages PDF
Abstract

ObjectivesAn LC-MS/MS method was developed for simultaneous quantitation of tacrolimus, sirolimus and everolimus in whole blood, and compared to HPLC-UV and immunoassay methods.Design and methodsBlood (0.1 mL) was analysed following solid-phase extraction and chromatographic resolution using a C18 column (45 °C) and mobile phase of methanol/40 mM ammonium acetate/glacial acetic acid (83/17/0.1) at 200 μL/min, with positive electrospray ionisation and multiple reaction monitoring.ResultsIntra- and inter-day imprecision and inaccuracy were ≤ 12.2% over a 1.5–40 μg/L calibration range. An external quality assurance programme confirmed acceptable inaccuracy and imprecision of the LC-MS/MS method, but highlighted problems with immunoassay quantitation, particularly for everolimus, showing a > 30% bias in FPIA everolimus concentrations measured in pooled patient samples versus spiked drug-free whole blood.ConclusionsLC-MS/MS provides significant accuracy and precision advantages compared to HPLC and immunoassays. Discrepancies in everolimus concentrations measured by the Seradyn FPIA immunoassay require further investigation.

Related Topics
Life Sciences Biochemistry, Genetics and Molecular Biology Biochemistry
Authors
, , ,