Article ID Journal Published Year Pages File Type
1969850 Clinical Biochemistry 2011 4 Pages PDF
Abstract

ObjectivesA new electrochemiluminescence testosterone immunoassay (Elecsys® Testo II) was evaluated.Design and methodsWithin-run precision, total precision, and method comparisons to the former immunoassay and LC-MS/MS were investigated. Reference intervals were established for adults and children.ResultsImprecision ranged from 0.9% to 8.5%. Method comparisons revealed slopes from 0.68 to 1.32 and coefficients of correlation from 0.893 to 0.991.ConclusionsImproved agreement with LC-MS/MS was demonstrated, especially for female patients. The assay may be a suitable alternative for laboratories with no access to MS techniques.

Related Topics
Life Sciences Biochemistry, Genetics and Molecular Biology Biochemistry
Authors
, , , , , , , , , , ,