Abstract

Objectives:To evaluate the inter-laboratory reproducibility of blood test for liver fibrosis: FibroMeter, Fibrotest, APRI and their composites variables.Design and methods:Four studies, including 147 patients, were performed: study #1 included 2 metachronous blood samples and 2 laboratories; studies #2, #3 and #4 included synchronous samples with assays delayed at day 1 in 12 laboratories, at day 0 in 10 laboratories and at day 0 or 1 in 2 laboratories, respectively. Agreement was evaluated by the intraclass correlation coefficient (ric).Results:In studies #1, #2 and #4, ric for FibroMeter was 0.893, 0.942 and 0.991, respectively. In study #3, the ric were: FibroMeter: 0.963, Fibrotest: 0.984, APRI: 0.949. Large simulated variations in composite variables had a weak impact on FibroMeter.Conclusions:When blood marker limits are controlled, inter-laboratory agreement of blood tests is excellent in clinical practice conditions. Blood tests are robust against the variability of composite blood variables.

Related Topics
Life Sciences Biochemistry, Genetics and Molecular Biology Biochemistry