Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1993319 | Methods | 2015 | 7 Pages |
A new vessel for simultaneous isothermal calorimetry and respirometry (calorespirometry) on terrestrial (non-aqueous) samples has been developed. All types of small (<1 g) biological samples (insects, soil, leaves, fungi, etc.) can be studied. The respirometric measurements are made by opening and closing a valve to a vial inside the sample ampoule containing a carbon dioxide absorbent. Typically a 7 h measurement results in seven measurements of heat production rate, oxygen consumption and carbon dioxide production, which can be used to evaluate how the metabolic activity in a sample changes over time. Results from three experiments on leaves, a cut vegetable, and mold are given. As uncertainties – especially in the carbon dioxide production – tend to be quite high, improvements to the technique are also discussed.