Article ID Journal Published Year Pages File Type
2859892 The American Journal of Cardiology 2008 5 Pages PDF
Abstract
The transcatheter closure of certain types of patent ductus arteriosus (PDA) remains a challenge. The investigators report initial clinical experience with 25 patients who underwent PDA occlusion with the use of a swivel-disk device (SDD) or a plug occluder (PO). The patients were divided into 2 groups: in the SDD group, 12 patients aged 1 to 2 years with type A PDA underwent attempted closure using the SDD, a modified Amplatzer duct occluder with a very low profile retention disk that can adapt itself at different PDA insertion angles. The mean PDA diameter was 4.8 ± 1.5 mm (range 3.8 to 8). In the PO group, 13 patients aged 0.5 to 3 years with type C, D, or E PDA underwent attempted occlusion with the PO, a tubular occluder made of Nitinol wire mesh. The mean PDA diameter was 4.2 ± 3.5 mm (range 1.2 to 9). The 2 occluders are filled with Dacron patches. The mean device diameters were 6.5 ± 1.2 and 6.8 ± 2.2 mm (range 4 to 11) in the SDD and PO groups, respectively. Complete angiographic closure was seen in 24 of 25 patients. The deployment of 4 5-PDA5 coils abolished a residual shunt associated with hemolysis in a 5-month-old patient with a large type C PDA after the implantation of an 11-mm PO. No other complications were observed. In conclusion, the SDD and the PO are promising additions to the armamentarium for PDA closure.
Related Topics
Health Sciences Medicine and Dentistry Cardiology and Cardiovascular Medicine
Authors
, , , , , ,