Article ID Journal Published Year Pages File Type
2943146 Journal of the American College of Cardiology 2015 11 Pages PDF
Abstract

Despite significant advances in battery longevity, lead performance, and programming features since the first implanted permanent pacemaker was developed, the basic design of cardiac pacemakers has remained relatively unchanged over the past 50 years. Because of inherent limitations in their design, conventional (transvenous) pacemakers are prone to multiple potential short- and long-term complications. Accordingly, there has been intense interest in a system able to provide the symptomatic and potentially lifesaving therapies of cardiac pacemakers while mitigating many of the risks associated with their weakest link—the transvenous lead. Leadless cardiac pacing represents the future of cardiac pacing systems, similar to the transition that occurred from the use of epicardial pacing systems to the familiar transvenous systems of today. This review summarizes the current evidence and potential benefits of leadless pacing systems, which are either commercially available (in Europe) or under clinical investigation.

Related Topics
Health Sciences Medicine and Dentistry Cardiology and Cardiovascular Medicine
Authors
, , , ,