Article ID Journal Published Year Pages File Type
2967641 Journal of Electrocardiology 2009 6 Pages PDF
Abstract

The AngelMed Guardian is an implantable medical device that records cardiac data and detects ischemic events using a standard pacemaker intracardiac lead positioned in the right ventricular apex. The Guardian has been implanted in 55 people in the United States and Brazil and is currently undergoing a Food and Drug Administration phase 2 pivotal trial in the United States. The Guardian detects acute ischemic events by analyzing ST-segment shifts. The ST-segment shifts are calculated as the difference between the ST deviation of a current 10-second electrogram window and a baseline ST deviation value. If the ST-segment shift is greater than a heart rate–dependent programmable threshold, then the device generates an emergency alert signal. Results thus far have demonstrated that (i) the intracardiac electrogram is relatively noise-free and (ii) the ST-shift technique used by the Guardian is effective for detecting acute ischemic events.

Related Topics
Health Sciences Medicine and Dentistry Cardiology and Cardiovascular Medicine
Authors
, , , , , ,