Article ID Journal Published Year Pages File Type
2969305 Journal of Electrocardiology 2008 8 Pages PDF
Abstract

AimThe aim of the study was to simultaneously test the EASI lead system and two other derived ECG methods against the standard 12-lead ECG during percutaneous coronary intervention (PCI).MethodsDuring 44 percutaneous coronary interventions, a simultaneously recorded 12-lead and EASI ECG were marked at the start of the PCI (baseline) and at known ischemia caused by balloon inflation (peak). ST deviations were measured 60 ms after the J point at baseline and peak in all leads and were summated (SUMST) to assess overall changes. For regional changes, the lead with the highest ST deviation (PEAKST) was marked. For each patient, derived 12-lead ECGs were computed from the EASI leads and a lead subset using patient-specific coefficients (PS) and coefficients based on a patient population (GEN). Absolute differences were computed between each derived and routine ECG for SUMST and PEAKST.ResultsSUMST was at baseline 567 μV (range: 150-1707) and increased at peak to 871 μV (range: 350-2101). SUMST difference at peak was for EASI: 163 μV (CI: 90-236, P <.001), GEN: 46 μV (CI: 2-91, P = .40), and PS: 16 μV (CI: 3-30, P = .15). PEAKST difference at peak was for EASI: 49 μV (CI: 19-220, P = .02), GEN: 48 μV (CI: -43-154, P = .26), and PS: 20 μV (CI: -51-32, P = .65).ConclusionSimultaneous direct comparison of three derived ECG methods shows overall and regional differences in accuracy across PS, GEN, and EASI. Median SUMST and PEAKST differences for PS are lower than for GEN and EASI, and show a more accurate reconstruction.

Related Topics
Health Sciences Medicine and Dentistry Cardiology and Cardiovascular Medicine