Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
3015531 | Revista Española de Cardiología | 2010 | 5 Pages |
Abstract
Currently, nuclear magnetic resonance imaging is contraindicated in patients with a pacemaker or implantable cardioverter-defibrillator. This study was carried out because the potential risks in this situation need to be clearly defined. This prospective study evaluated clinical and electrical parameters before and after magnetic resonance imaging was performed in 33 patients (five with implantable cardioverter-defibrillators and 28 with pacemakers). In these patients, magnetic resonance imaging was considered clinically essential. There were no clinical complications. There was a temporary communication failure in two cases, sensing errors during imaging in two cases, and a safety signal was generated in one pacemaker at the maximum magnetic resonance frequency and output level. There were no technical restrictions on imaging nor were there any permanent changes in the performance of the cardiac pacing device.
Related Topics
Health Sciences
Medicine and Dentistry
Cardiology and Cardiovascular Medicine
Authors
Francisco BuendÃa, Juan M. Sánchez-Gómez, MarÃa J. Sancho-Tello, José Olagüe, JoaquÃn Osca, Ãscar Cano, Miguel A. Arnau, Begoña Igual,