Article ID Journal Published Year Pages File Type
4241715 Journal of Vascular and Interventional Radiology 2007 4 Pages PDF
Abstract
The feasibility of imaging pulsatility in an aneurysm model with the high-resolution dynamic multiscan technique of 64-slice computed tomography (CT) was studied. A pulsatile aneurysm phantom was constructed and imaged with dynamic multiscan technique. The aneurysm model was filled with iodinated contrast material (250 Hounsfield Units) and was scanned with use of a gantry rotation time of 0.33 seconds, slice thickness of 1.2 mm, effective coverage of 24 mm, and total imaging time of 4 seconds. Images were reconstructed at 50-msec intervals. The visualization of wall motion was qualitatively evaluated by direct comparison of four-dimensional images versus phantom motion. Pulsatility imaging without perceptible artifact or need for cardiac gating was achieved with the use of this technique.
Related Topics
Health Sciences Medicine and Dentistry Radiology and Imaging
Authors
, , , , , , ,