Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
4964867 | Computers in Biology and Medicine | 2017 | 32 Pages |
Abstract
Functional Near Infrared Spectroscopy (fNIRS) is an emerging neuroimaging technique that utilizes near infrared light to detect cortical concentration changes of oxy-hemoglobin and deoxy-hemoglobin non-invasively. Using light sources and detectors over the scalp, multi-wavelength light intensities are recorded as time series and converted to concentration changes of hemoglobin via modified Beer-Lambert law. Here, we describe a potential source for systematic error in the calculation of hemoglobin changes and light intensity measurements. Previous system characterization and analysis studies looked into various fNIRS parameters such as type of light source, number and selection of wavelengths, distance between light source and detector. In this study, we have analyzed the contribution of light detector surface area to the overall outcome. Results from Monte Carlo based digital phantoms indicated that selection of detector area is a critical system parameter in minimizing the error in concentration calculations. The findings here can guide the design of future fNIRS sensors.
Related Topics
Physical Sciences and Engineering
Computer Science
Computer Science Applications
Authors
Lei Wang, Hasan Ayaz, Meltem Izzetoglu, Banu Onaral,