Article ID Journal Published Year Pages File Type
505197 Computers in Biology and Medicine 2012 12 Pages PDF
Abstract

This paper presents a new index to measure the hypnotic depth of anaesthesia (DoA) using EEG signals. This index is derived from applying combined Wavelet transform, eigenvector and normalisation techniques. The eigenvector method is first applied to build a feature function for six levels of coefficients in a discrete wavelet transform (DWT). The best Daubechies wavelet and their ranking value p are optimally determined to identify different states of anaesthesia. A statistic normalisation process is then carried out to re-scale data and compute the hypnotic depth of anaesthesia. Finally, a new function ZDoA is proposed to compute a DoA index which corresponds one of the five depths of anaesthesia states to very deep anaesthesia, deep anaesthesia, moderate anaesthesia, light anaesthesia and awake. Simulation results based on real anaesthetised EEGs demonstrate that the new index generally parallels the BIS index. In particular, the ZDoA index is often faster than the BIS index to react to the transition period between consciousness and unconsciousness for this data set. A Bland–Altman plot indicates a 95.23% agreement between the ZDoA and BIS indices. The ZDoA trend is responsive, and its movement is consistent with the clinically observed and recorded changes of the patients.

Related Topics
Physical Sciences and Engineering Computer Science Computer Science Applications
Authors
, , , ,