Article ID Journal Published Year Pages File Type
505576 Computers in Biology and Medicine 2009 10 Pages PDF
Abstract

During ambulatory monitoring, it is often required to record the electroencephalogram (EEG) and the electrocardiogram (ECG) simultaneously. It would be ideal if both EEG and ECG can be obtained with one measurement. We introduce an algorithm combining the wavelet shrinkage and signal averaging techniques to extract the EEG and ECG components from an EEG lead signal to a noncephalic reference (NCR). The evaluation using simulation data and measured data showed that the normalized power spectrum unvaried in all frequency bands for the EEG components, and the sensitivity and specificity of R-wave detection for the ECG component were nearly 100%.

Related Topics
Physical Sciences and Engineering Computer Science Computer Science Applications