Article ID Journal Published Year Pages File Type
505587 Computers in Biology and Medicine 2008 15 Pages PDF
Abstract

Finite element model of radiofrequency ablation (RFA) with cooled-tip probe in liver has been developed by employing COMSOL Multiphysics software. It describes coupled electric, thermal and sodium chloride solution infiltration flow phenomena taking place during ablation processes. Features of hydraulic capacity, saturation of the tissue by infiltration, and dependency of electrical conductivity on the damage integral of the tissue have been supplied to the model. RFA experiments have validated the model. Physical parameters describing hydraulic capacity and hydraulic conductivity in the tissue, as well as, the relation of electrical conductivity against the value of damage integral have been determined.

Related Topics
Physical Sciences and Engineering Computer Science Computer Science Applications
Authors
, , , ,