Article ID Journal Published Year Pages File Type
5131171 Analytica Chimica Acta 2017 7 Pages PDF
Abstract

•An easily and fast procedure to produce sealed μPADs is proposed.•μPADs are fabricated using low-cost materials and equipment.•Hundreds of 2D and 3D μPADs can be produces rapidly.•μPADs have been successfully applied in blood and urine analysis.

This paper describes the fabrication of 2D and 3D microfluidic paper-based analytical devices (μPADs) for monitoring glucose, total protein, and nitrite in blood serum and artificial urine. A new method of cutting and sealing filter paper to construct μPADs was demonstrated. Using an inexpensive home cutter printer soft cellulose-based filter paper was easily and precisely cut to produce pattern hydrophilic microchannels. 2D and 3D μPADs were designed with three detection zones each for the colorimetric detection of the analytes. A small volume of samples was added to the μPADs, which was photographed after 15 min using a digital camera. Both μPADs presented an excellent analytical performance for all analytes. The 2D device was applied in artificial urine samples and reached limits of detection (LODs) of 0.54 mM, 5.19 μM, and 2.34 μM for glucose, protein, and nitrite, respectively. The corresponding LODs of the 3D device applied for detecting the same analytes in artificial blood serum were 0.44 mM, 1.26 μM, and 4.35 μM.

Graphical abstractDownload high-res image (262KB)Download full-size image

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , ,