Article ID Journal Published Year Pages File Type
5396892 Journal of Electron Spectroscopy and Related Phenomena 2008 12 Pages PDF
Abstract

Using the basic spectrometer trajectory equation for motion in an ideal 1/r potential derived in Eq. (101) of part I [T.J.M. Zouros, E.P. Benis, J. Electron Spectrosc. Relat. Phenom. 125 (2002) 221], the operational characteristics of a hemispherical deflector analyser (HDA) such as dispersion, energy resolution, energy calibration, input lens magnification and energy acceptance window are investigated from first principles. These characteristics are studied as a function of the entry point R0 and the nominal value of the potential V(R0) at entry. Electron-optics simulations and actual laboratory measurements are compared to our theoretical results for an ideal biased paracentric HDA using a four-element zoom lens and a two-dimensional position sensitive detector (2D-PSD). These results should be of particular interest to users of modern HDAs utilizing a PSD.

Related Topics
Physical Sciences and Engineering Chemistry Physical and Theoretical Chemistry
Authors
, ,