Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
5499049 | Radiation Physics and Chemistry | 2017 | 5 Pages |
Abstract
In a report detailing an international comparison of air kerma standards for the ISO 4037 narrow spectrum series (EUROMET.RI(I)-S3), which was conducted from 2004 to 2005, the ratio of differences and expanded uncertainties (Di/Ui) for INER's IN-250 and IN-300 radiation qualities were 0.9 and 1.8, respectively. If the correction factors obtained in this study are substituted, the differences can be reduced, and the (Di/Ui) ratios become 0.36 and 0.6 for IN-250 and IN-300, respectively.
Keywords
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Radiation
Authors
Tseng-Te Huang, Chien-Hau Chu, Yi-Chun Lin,