Dosimetric properties of radiation beams used in radiotherapy are directly related to the energy spectrum produced by the treatment unit. Therefore, the development of methodologies to evaluate in a simple and accurate way the spectra of clinical beams can help establishing the quality control of the treatment. The purpose of this study is to present a practical and low cost methodology for determining primary spectra of radiotherapy photon beams from transmission measurements in attenuators of aluminum and using the method of the inverse Laplace transform. Monte Carlo simulation with PENELOPE code was used in order to evaluate and validate the reconstructed spectra by the calculation of dosimetric parameters that characterize the beam. Percentage depth dose values simulated with a 6Â MV reconstructed spectrum shows maximum difference of 4.4% when compared to values measured at the corresponding clinical beam. For a 10Â MV beam that difference was around 4.2%. Results obtained in this study confirm the adequacy of the proposed methodology for assessing primary photon beams produced by clinical accelerators.