Abstract
All bead types showed TL to be linear with dose (R2>0.999). Glow curve dosimetric peaks reached a maximum value at 300 °C for the Toho and 290 °C for the Czech and Mill Hill products but was between 200-250 °C for the Rocaille product. Radiation sensitivity following mass normalisation varied within an order of magnitude; Toho samples showed the greatest and Rocaille the least sensitivity. For the Toho, Czech, Rocaille and Mill Hill samples the energy responses at 80 kVp were 5.0, 4.0, 3.6 and 3.3 times that obtained at 6 MV. All four glass bead types offer potential use as TL dosimeters over doses commonly applied in radiotherapy. Energy response variation was <1% at 6 MV but significant variation was found for photon beam energies covering the kV range; careful characterisation is required if use at this range is intended.