Article ID Journal Published Year Pages File Type
557647 Biomedical Signal Processing and Control 2010 9 Pages PDF
Abstract

Patients with type 1 diabetes require insulin therapy to maintain blood glucose levels within safe ranges since their pancreas is unable to complete its function. The development of a closed-loop artificial pancreas capable of maintaining normoglycemia during daily life will dramatically improve the quality of life for insulin-dependent diabetic patients. In this work, a closed-loop control strategy for blood glucose level regulation in type 1 diabetic patients is presented. A robust controller is designed using a combination of internal model and sliding mode control techniques. Also, the controller is provided with a feedforward loop to improve meal compensation. A simulation environment designed for testing the artificial pancreas control algorithms has been used to evaluate the controller. The simulation results show a good controller performance in fasting conditions and meal disturbance rejection, and robustness against model–patient mismatch and meal estimation errors.

Related Topics
Physical Sciences and Engineering Computer Science Signal Processing
Authors
, , ,