Article ID Journal Published Year Pages File Type
5930948 The American Journal of Cardiology 2013 7 Pages PDF
Abstract
Quantitative assessment of aortic regurgitation (AR) remains challenging. The present study evaluated the accuracy of 2-dimensional (2D) and 3-dimensional (3D) transthoracic echocardiography (TTE) for AR quantification, using 3D 3-directional velocity-encoded magnetic resonance imaging (VE-MRI) as the reference method. Thirty-two AR patients were included. With color Doppler TTE, 2D effective regurgitant orifice area (EROA) was calculated using the proximal isovelocity surface area method. From the 3D TTE multiplanar reformation data, 3D-EROA was calculated by planimetry of the vena contracta. Regurgitant volumes (RVol) were obtained by multiplying the 2D-EROA and 3D-EROA by the velocity-time integral of AR jet and compared with that obtained using VE-MRI. For the entire population, 3D TTE RVol demonstrated a strong correlation and good agreement with VE-MRI RVol (r = 0.94 and −13.6 to 15.6 ml/beat, respectively), whereas 2D TTE RVol showed a modest correlation and large limits of agreement with VE-MRI (r = 0.70 and −22.2 to 32.8 ml/beat, respectively). Eccentric jets were noted in 16 patients (50%). In these patients, 3D TTE demonstrated an excellent correlation (r = 0.95) with VE-MRI, a small bias (0.1 ml/beat) and narrow limits of agreement (−18.7 to 18.8 ml/beat). Finally, the kappa agreement between 3D TTE and VE-MRI for grading of AR severity was good (k = 0.96), whereas the kappa agreement between 2D TTE and VE-MRI was suboptimal (k = 0.53). In conclusion, AR RVol quantification using 3D TTE is accurate, and its advantage over 2D TTE is particularly evident in patients with eccentric jets.
Related Topics
Health Sciences Medicine and Dentistry Cardiology and Cardiovascular Medicine
Authors
, , , , , , , , , , , , , , , , , , , , ,