Article ID Journal Published Year Pages File Type
6489774 Journal of Bioscience and Bioengineering 2018 7 Pages PDF
Abstract
Portable, on-site blood typing methods will help provide life-saving blood transfusions to patients during an emergency or natural calamity, such as significant earthquakes. We have previously developed waveguide-mode (WM) sensors for forward ABO and Rh(D) blood typing and detection of antibodies against hepatitis B virus and hepatitis C virus. In this study, we evaluated a WM-sensor for reverse ABO blood typing. Since reverse ABO blood typing is a method for detection of antibodies against type A and type B oligosaccharide antigens on the surface of red blood cells (RBCs), we fixed a synthetic type A or type B trisaccharide antigen on the sensor chip of the WM sensor. We obtained significant changes in the reflectance spectra from a WM sensor on type A antigen with type B plasma and type O plasma and on type B antigen with type A plasma and type O plasma, and no spectrum changes on type A antigen or type B antigen with type AB plasma. Signal enhancement with the addition of a peroxidase reaction failed to increase the sensitivity for detection on oligosaccharide chips. By utilizing hemagglutination detection using regent type A and type B RBCs, we successfully determined reverse ABO blood groups with higher sensitivity compared to a method using oligosaccharide antigens. Thus, functionality of a portable device utilizing a WM sensor can be expanded to include reverse ABO blood typing and, in combination with forward ABO typing and antivirus antibody detection, may be useful for on-site blood testing in emergency settings.
Related Topics
Physical Sciences and Engineering Chemical Engineering Bioengineering
Authors
, , , , , , , , ,