We present an FTIR based assay to monitor the whole cell mediated oxidation of aryl alcohols by measuring the characteristic IR absorption of the hydroxyl group [OH] of the substrate and the carbonyl group [CO] of the corresponding oxidized product. This method expedites the analysis of whole cell mediated catalysis which is usually done by GC and/or HPLC. The FTIR assay had linearity with R2 â¥Â 0.980 and sensitivity up to 10 μM. The accuracy and precision of FTIR assay was found â¥81% and â¥94%, respectively. This assay was validated by GC which exhibited â¥82% accuracy and â¥79% precision. The time of analysis taken by this assay was 2-3 min per sample in comparison with 20-40 min by GC.
Related Topics
Physical Sciences and Engineering
Chemical Engineering
Bioengineering
Authors
Sneha Sudhakara, Anju Chadha,