Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
6891056 | Computer Methods and Programs in Biomedicine | 2018 | 27 Pages |
Abstract
The test results were evaluated separately in terms of measurement accuracy, working time, the continuity of the Bluetooth connection and the reliability of the measurement results. The measurement accuracy for systolic, diastolic blood pressure and heart rate values was obtained as 93.89%, 91.95% and 97.66%, respectively. The maximum number of measurements which can be conducted with four 1000â¯mA alkaline batteries at 20 min intervals was found approximately 79 (little more than 24 h). In addition, it was determined that the continuity of the Bluetooth connection and the reliability of the measurement results were automatically achieved through the features in the interface developed for the smart mobile device. These different evaluations showed that the smart blood pressure holter can meet the requirements of a classical holter device.
Keywords
Related Topics
Physical Sciences and Engineering
Computer Science
Computer Science (General)
Authors
İlhan İlhan,