Abstract
This paper presents a multi-model robust control scheme to control the depth of hypnosis during intravenous administration of propofol. The objective of the proposed control scheme is to provide an adequate drug administration regime for propofol to avoid overdosing and underdosing of patients. The proposed scheme is designed to withstand the patient's inherent drug response variability, to achieve good output disturbance and sensor noise rejection, and to attain a good set point response. A comprehensive simulation study of 44 patients is presented to assess the performance of the proposed control scheme.