Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
7141728 | Sensors and Actuators B: Chemical | 2018 | 20 Pages |
Abstract
The flow injection analysis (FIA) methodology here reported for the first time for acetaldehyde analysis in cider, allowed to obtain results not statistically different from the ones obtained with Gas Chromatography-Flame Ionization Detection (GC-FID) (α = 0.05). Besides, the new methodology does not use expensive instrumentation and the analysis time is similar. Detection limit of this new method was 2.0 mg Lâ1, and quantification limit was 6.7 mg Lâ1, not much higher than the values of 1.4 mg Lâ1 and 4.6 mg Lâ1, respectively, of the GC-FID.
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
Marta I.S. VerÃssimo, José A.F. Gamelas, Mário M.Q. Simões, Dmitry V. Evtuguin, M. Teresa S.R. Gomes,