Abstract
The performance of the corresponding imprinted and non-imprinted electrodes was evaluated by EIS and the effect of several variables, such as monomer and template concentrations, or thickness of imprinting surface, was controlled and optimized by the number of CV cycles. The detection limit of the MIP-based sensors was 0.60 nM in MES buffer. High repeatability and good selectivity were observed in the presence of a model protein BSA. The sensor performance was also tested to check the effect of inorganic ions in tap water. The detection limit observed was 16.83 nM, with a recovery factor of 91.1 ± 6.6%. The sensor described in this work is a potential tool for screening PA on-site, due to the simplicity of fabrication, disposability, short response time, low cost, good sensitivity and selectivity.