Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
7149999 | Sensors and Actuators B: Chemical | 2012 | 6 Pages |
Abstract
An all-solid-state sensor based on screen-printing technique for rapid blood calcium test was studied. The sensor consists of a thick film calcium ion-selective electrode and a pseudo-reference electrode. Poly(3,4-ethylenedioxythiophene) (PEDOT) doped with poly(styrenesulfonate) (PSS) was screen-printed on carbon paste electrodes as internal solid contacts for both the ion-selective and reference electrodes. PVC membranes with and without neutral carrier ETH129 were coated on the PEDOT (PSS) layer to form calcium ion-selective and pseudo-reference electrodes, respectively. The properties of screen-printed PEDOT (PSS) film on carbon paste electrode were studied by electrochemical impedance spectroscopy (EIS) and cyclic voltammetry (CV) methods. The response time of the calcium sensors was 15 s while less than 15 μL sample was applied. And the sensitivity was 30.0 ± 1.0 mV per decade, while the potentiometric log selectivity coefficients were â3.80, â3.35 and â4.60 for Na+, K+ and Mg2+. Other properties of the calcium sensors were evaluated, including the temperature effect, repeatability, reproducibility and stability. The testing results with the calcium sensors were compared with those with a routine hospital electrolyte analyzer. The clinical trial results reveal that the all-solid-state calcium sensor is promising for the point of care testing.
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
You Wang, Hui Xu, Xiaoxi Yang, Zhiyuan Luo, Jianming Zhang, Guang Li,