Article ID Journal Published Year Pages File Type
743244 Sensors and Actuators B: Chemical 2009 9 Pages PDF
Abstract

The conductance of several tin oxide gas sensitive layers was simultaneously measured in thermo-cyclic and isothermal operation mode at various concentrations of CO and propene in air, respectively. Different measurement conditions were set by the cycle time, the gas flow rate, the humidity and the thickness of the sensitive layer. The sensor response is expressed by the conductance-over-time profiles (CTPs), the gas sensitivity or by the sum of the CTP sampling points. The resultant sensitivities from the CTPs were found to be higher than those of the isothermal measurements. The CTPs considerably change with the kind of gas as well as with the variation of the measurement conditions. This is discussed for both types of target gases with respect to the temperature, the adsorbates assumed and the reaction kinetics in relation to the gas transport conditions in the porous gas sensitive layer.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry