Article ID Journal Published Year Pages File Type
743978 Sensors and Actuators B: Chemical 2011 8 Pages PDF
Abstract

We have investigated real-time, label-free, in situ detection of human epidermal growth factor receptor 2 (Her2) in diluted serum using the first longitudinal extension mode of a lead zirconate-lead titanate (PZT)/glass piezoelectric microcantilever sensor (PEMS) with H3 single-chain variable fragment (scFv) immobilized on the 3-mercaptopropyltrimethoxysilane (MPS) insulation layer of the PEMS surface. We showed that with the longitudinal extension mode, the PZT/glass PEMS consisting of a 1 mm long and 127 μm thick PZT layer bonded with a 75 μm thick glass layer with a 1.8 mm long glass tip could detect Her2 at a concentration of 6–60 ng/ml (or 0.06–0.6 nM) in diluted human serum, about 100 times lower than the concentration limit obtained using the lower-frequency flexural mode of a similar PZT/glass PEMS. We further showed that with the longitudinal mode, the PZT/glass PEMS determined the equilibrium H3–Her2 dissociation constant Kd to be 3.3 ± 0.3 × 10−8 M consistent with the value, 3.2 ± 0.28 × 10−8 M deduced by the surface plasmon resonance method (BIAcore).

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , ,