Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
745401 | Sensors and Actuators B: Chemical | 2006 | 6 Pages |
Abstract
A cobaloxime ([chlorobis(dimethylglyoximeato)(triphenylphosphine)] cobalt (III), [Co(dmgH)2pph3Cl]) incorporated in a plasticized poly(vinyl chloride) membrane was used to develop a perchlorate-selective electrode. The influence of membrane composition on the electrode response was studied. The electrode exhibits a Nernstian response over the perchlorate concentration range 1.0 × 10−6 to 1 × 10−1 mol l−1 with a slope of −56.8 ± 0.7 mV per decade of concentration, a detection limit of 8.3 × 10−7, a wide working pH range (3–10) and a fast response time (<15 s). The electrode shows excellent selectivity towards perchlorate with respect to many common anions. The electrode was used to determine perchlorate in water and human urine.
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
M.A. Zanjanchi, M. Arvand, M. Akbari, K. Tabatabaeian, G. Zaraei,