Article ID Journal Published Year Pages File Type
750729 Sensors and Actuators B: Chemical 2015 6 Pages PDF
Abstract

The development of a sandwich-type biosensor for glucose quantification is presented. This work is focused on the optimization of the enzymatic matrix of the biosensor. The best performance was found for an enzymatic matrix composed by 30% w/w mucin, 70% w/w albumin, 1.35 U glucose oxidase (GOX) per sensor, and glutaraldehyde diluted to 3%. The crosslinking with glutaraldehyde transforms this mixture into a hydrogel that is entrapped between two membranes of polycarbonate. The selected sandwich-type biosensor showed very good response time, sensitivity, stability, and sensor-to-sensor reproducibility.According to the results presented in this manuscript, a biosensor prepared with very high amount of enzyme would not necessarily increase the analytical signal. Simulated curves are compared with experimental data to explain the dependence of sensitivity on the concentration of enzyme. In addition, this kind of comparison represents a quite simple way to estimate the value of vmax ≈0.13 M s−1 from the amperometric response of a sensor prepared with 1.34 U of GOX.Considering that sandwich-type biosensors are commonly assembled as part of devices where the sample is diluted with buffer, the more than 3 orders of magnitude of linear behavior of this sensor would ensure the possibility for assessing any sample.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , ,