Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
8171694 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2015 | 4 Pages |
Abstract
In this work we present the performance of a fast acquisition system for in-beam PET monitoring during the irradiation of a PMMA phantom with a clinical proton beam. The experimental set-up was based on 4 independent detection modules. Two detection modules were placed at one side of a PMMA phantom and the other two modules were placed at the opposite side of the phantom. One detection module was composed of a Silicon Photon Multiplier produced by AdvanSiD coupled to a single scintillating LYSO crystal. The read-out system was based on the TOFPET ASIC managed by a Xilinx ML605 FPGA Evaluation Board (Virtex 6). The irradiation of the PMMA phantom was performed at the CNAO hadrontherapy facility (Pavia, Italy) with a 95 MeV pulsed proton beam. The pulsed time structure of the proton beam was reconstructed by each detection module. The β+ annihilation peak was successfully measured and the production of β+ isotopes emitters was observed as increasing number of 511 keV events detected during irradiation. Finally, after the irradiation, the half lives of the 11C and 15O radioactive isotopes were estimated.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
M.A. Piliero, M.G. Bisogni, P. Cerello, A. Del Guerra, E. Fiorina, B. Liu, M. Morrocchi, F. Pennazio, G. Pirrone, R. Wheadon,