Abstract
This study describes the development of a CdTe pixel detector for high-energy X-ray diffraction experiments at synchrotron radiation facilities. We have developed three prototypes. The third prototype detector (SP8-02B) was designed with a pixel size of 200μmÃ200μm and a matrix of 20Ã50 pixels. The sensor was bonded to an application-specific integrated circuit with a preamplifier, shaper, window-type discriminator, and a 20-bit counter for each pixel. The SP8-02B detector improved the uniformity between pixels in one chip in comparison with the previous prototypes. The long-term stability was measured at various temperatures. A moderate cooling achieved good stability that was not affected by the polarization effect.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
T. Hirono, H. Toyokawa, Y. Furukawa, M. Kawase, T. Ohata, S. Wu, H. Ikeda, G. Sato, T. Takahashi, S. Watanabe,