Article ID Journal Published Year Pages File Type
8179474 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2013 4 Pages PDF
Abstract
The γ-ray spectroscopy is a promising method for diagnosing fast ions and confined α particles in a fusion plasma device. This application requires γ-ray detectors with high energy resolution (say a few percent for γ-ray energies in the range 1-5 MeV), high efficiency and high count rate capability, ideally up to a few MHz. Furthermore, the detector will have to withstand the high 14 MeV and 2.45 MeV neutron fluxes produced by the main fusion reactions between deuterium and tritium. Experimental results demonstrate that the requirements on energy resolution, efficiency and count rate can be met with a LaBr3(Ce) scintillator detector equipped with fast digital data acquisition. The measured response of the detector to 2.45 MeV neutrons is presented in this paper and discussed in terms of the interaction mechanism between neutrons and detector.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation
Authors
, , , , , , , , ,