Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
8208771 | Applied Radiation and Isotopes | 2018 | 13 Pages |
Abstract
A comparison of calculations of the activity of a 3H2O liquid scintillation source using the same experimental data set collected at the LNE-LNHB with a triple-to-double coincidence ratio (TDCR) counter was completed. A total of 17 laboratories calculated the activity and standard uncertainty of the LS source using the files with experimental data provided by the LNE-LNHB. The results as well as relevant information on the computation techniques are presented and analysed in this paper. All results are compatible, even if there is a significant dispersion between the reported uncertainties. An output of this comparison is the estimation of the dispersion of TDCR measurement results when measurement conditions are well defined.
Keywords
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Radiation
Authors
Philippe Cassette, Timotheos Altzitzoglou, Andrei Antohe, Mario Rossi, Arzu Arinc, Marco Capogni, Raphael Galea, Arunas Gudelis, Karsten Kossert, K.B. Lee, Juncheng Liang, Youcef Nedjadi, Pilar Oropesa Verdecia, Tanya Shilnikova, Winifred van Wyngaardt,