Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
8252197 | Radiation Physics and Chemistry | 2016 | 6 Pages |
Abstract
Due to the necessity of using low background NaI detector in sensitive and accurate measurements, study on the response function variations in different conditions is very important. These types of detectors have different responses in various measurement conditions, including time, temperature and high voltage. In this study, the response function of 76 B 76 LB NaI (SCIONIX) in different conditions is discussed. According to the channel shifting in these detectors and its direct effect on degrading the resolution, the most convenient measurement condition for these detectors, is proposed. Finally, it is recommended that before long-time measurements a “waiting time” is needed to avoid the channel shifting effects.
Keywords
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Radiation
Authors
Y. Rezaei Moghaddam, L. Rafat Motavalli, H. Miri Hakimabadi,