Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
8315685 | Clinica Chimica Acta | 2012 | 7 Pages |
Abstract
⺠Accuracy of a hand-held glucometer and a continuous monitoring device was evaluated. ⺠Overall, laboratory values did not differ from glucometer and CGM readings. ⺠Below the hypoglycemic threshold the error was > 0.8 mM in 78.9% and 94.1% of cases. ⺠The errors of the CGM device were inversely related to the rate of glucose change. ⺠Future portable meters performance should be focused on the extreme glucose levels.
Keywords
Related Topics
Life Sciences
Biochemistry, Genetics and Molecular Biology
Biochemistry
Authors
M.P. Francescato, M. Geat, G. Stel, S. Cauci,