Article ID Journal Published Year Pages File Type
8316919 Clinical Biochemistry 2018 17 Pages PDF
Abstract
We have evaluated the analytical performances of Cobas 8000® analyzer for the quantification of albumin and IgG in serum and CSF. Coefficients of variation were below 3.3% for within-run precision and below 3.1% for between-run precision. Results were similar or better than those obtained on nephelometer Immage 800® and turbidimeter SPAPLUS®. The uncertainty of quantification of IgG in CSF was 9% and that of albumin in CSF was 12%. IgG and albumin measured on Cobas 8000® in serum and CSF showed good agreement with results obtained on the nephelometer Immage 800®, including for the classification of albumin quotient and CSF IgG index as normal or pathological. Therefore, Cobas 8000® is a valuable tool for the quantification of IgG and albumin in CSF, in the context of diagnosis of MS and other inflammatory disease affecting the central nervous system.
Related Topics
Life Sciences Biochemistry, Genetics and Molecular Biology Biochemistry
Authors
,