Abstract
The new BRAHMS KRYPTOR sFlt-1 and PlGF immunoassay show excellent precision and reliability. The assay results and the diagnostic capacity were highly comparable to established fully automated immunoassays (Elecsys). Hence, sFlt-1/PlGF ratio generated on KRYPTOR immunoassay platform should be suitable for diagnosing preeclampsia in clinical routine laboratory.