Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
8317837 | Clinical Biochemistry | 2012 | 6 Pages |
Abstract
⺠We evaluated the MODULAR E170 and the ADVIA Centaur Total Vitamin D assays. ⺠Both assays demonstrated acceptable linearity, imprecision, and accuracy. ⺠The E170 assay demonstrated consistent underestimation of 25-OH vitamin D2 levels. ⺠The ADVIA Centaur assay demonstrated a smaller average bias than the E170 assay. ⺠The ADVIA Centaur assay needs to improve its analytical precision and interference.
Keywords
Related Topics
Life Sciences
Biochemistry, Genetics and Molecular Biology
Biochemistry
Authors
Yu Chen, Lois Kinney, Andrea BožoviÄ, Hilary Smith, Heather Tarr, Eleftherios P. Diamandis, Adrien LeBlanc,