Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
8335882 | International Journal of Biological Macromolecules | 2011 | 7 Pages |
Abstract
The prepared enzyme membrane was used for the construction of amperometric biosensor for urea detection. Its basic amperometric characteristics were investigated. A calibration plot was obtained for urea concentration ranging from 1.6 to 23 mM. A linear interval was detected along the calibration curve from 1.6 to 8.2 mM. The sensitivity of the constructed biosensor was calculated to be 3.1927 μA mMâ1 cmâ2. The correlation coefficient for this concentration range was 0.998. The detection limit with regard to urea was calculated to be 0.5 mM at a signal-to-noise ratio of 3. The biosensor was employed for 10 days while the maximum response to urea retained 86.8%.
Related Topics
Life Sciences
Biochemistry, Genetics and Molecular Biology
Biochemistry
Authors
Katya Gabrovska, Javor Ivanov, Ioana Vasileva, Nedyalka Dimova, Tzonka Godjevargova,