Abstract
Inhaled nitric monoxide (iNO) is increasingly used as a medical treatment for acute respiratory distress syndrome. A course of the existing nitric monoxide (NO) therapy with gas cylinders could cost up to approximately $15,000 for an average of 30.2Â h. Moreover, a gas cylinder containing a mixture of N2 and NO may potentially leak NO. The objective of this study is to develop an efficient and cost-effective on-site iNO generation system. In the present setup, NO was generated by using dry air or mixed oxygen/nitrogen (O2/N2) and an AC power source with an output power level of 5-30Â WÂ at atmospheric pressure. The simultaneously produced NO2 was eliminated with an ammonium sulfite ((NH4)2SO3) solution. The effects of the O2/N2 ratio, gas flow rate, discharge gap distance, output energy density and electrode structure on NOx concentration and the NO/NO2 ratio are reported. The concentrations of NO and NO2 reached 62Â ppm and 3Â ppm, respectively, after absorption and dilution at a gas flow rate of 6Â L/min. With the present setup, the AC arc discharge produced NOx at a stable concentration for at least 6Â h using dry air.