Article ID Journal Published Year Pages File Type
8851 Biomaterials 2010 8 Pages PDF
Abstract

Using a custom-built, implantable pulse generator, we studied the effects of small pulsed currents on the viability on rat aortic-derived cells (RAOC) in vitro. The pulsed currents (0.37 A/m2) underwent apoptosis within 24 h as shown by the positive staining for cleaved caspase-3 and classically apoptotic morphology. Based on these findings, we examined the effects of nanocurrents in vivo. The pulse generator was implanted subcutaneously in the rat model. The electrode|tissue interface histology revealed no difference between the active platinum surface and the neighboring control surface, however we found a large difference between electrodes that were functional during the entire experiment and non-active electrodes. These non-active electrodes showed an increase in impedance at higher frequencies 21 days post-implantation, whereas working electrodes retained their impedance value for the entire experiment. These results indicate that applied currents can reduce the impedance of implanted electrodes.

Related Topics
Physical Sciences and Engineering Chemical Engineering Bioengineering