Article ID Journal Published Year Pages File Type
9170169 Journal of Arrhythmia 2005 4 Pages PDF
Abstract
A 72-year-old man with a thin build had an ICD system with a generator implanted at left prepectoral space. The generator was exposed through thin overlying skin at 11 months following surgery. Although it was undermined with the adjacent skin, it was exposed again 6 months later. The generator was replaced in the ipsilateral subpectoral space. Since then, no signs of recurrence have been observed for the subsequent 12 months, with the patient pleased with its cosmetic appearance. This case illustrates the benefits of subpectoral implantation in the current ICD era in which subcutaneous implantation is common.
Related Topics
Health Sciences Medicine and Dentistry Cardiology and Cardiovascular Medicine
Authors
, , , , , , , ,