Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
9743409 | Analytica Chimica Acta | 2005 | 8 Pages |
Abstract
The paper presents a new conception of determination of inorganic speciation forms of arsenic: As(III) and As(V) as well selenium Se(IV) and Se(VI) by means of the high performance liquid chromatography hyphenated with a detection by the atomic absorption spectrometry with hydride generation (HPLC-HG-AAS). The application of optimization procedure conditions of chromatographic separation of arsenic and selenium speciation forms (using anion-exchange Supelco LC-SAX1 column and phosphate buffer at pH 5.40 as a mobile phase) as well as the use of the atomic absorption spectrometry as a detector, which enables work in fast sequential mode, allowed to develop original detection methodology of simultaneous determination of arsenic As(III), As(V) and selenium Se(IV) and Se(VI) speciation forms within a 220 s single analysis. The obtained detection limits were 7.8 ng mLâ1 for As(III); 12.0 ng mLâ1 for As(V); 2.4 ng mLâ1 for Se(IV) and 18.6 ng mLâ1 for Se(VI) and precision 10.5%, 12.1%, 14.2% and 17.3%, respectively, for 100 ng mLâ1. The described method was used for ground water analysis.
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
P. Niedzielski,