Ultrasonic slurry sampling electrothermal vaporization inductively coupled plasma mass spectrometry (USS-ETV-ICP-MS) has been applied to determine Ga, Ge, As, Se and Sb in fly ash samples. The influences of instrument operating conditions and slurry preparation on the ion signals were reported. Pd and ascorbic acid were used as the mixed modifiers to enhance the ion signals. This method has been applied to determine Ga, Ge, As, Se and Sb in NIST SRM 1633a and 1633b coal fly ash reference materials and a fly ash sample collected locally. Since the sensitivities of the elements studied in slurry solution and aqueous solution were different slightly, analyte addition technique was used for the determination of Ga, Ge, As, Se and Sb in these samples. The As and Se analysis results of the reference materials agreed with the certified values. The results for which no certified value was available were also found to be in good agreement between the ETV-ICP-MS results and the reference values. The reference value was obtained by digesting the samples and analyzing the digested sample solutions by pneumatic nebulization Dynamic Reaction Cell™ (DRC) ICP-MS. The method detection limits estimated from analyte addition curves were about 0.23, 0.13, 0.17, 0.25 and 0.11 μg g−1 for Ga, Ge, As, Se and Sb, respectively, in original fly ash samples.